Calculate sample to detector distance

Where 5.838 nm is the d001 spacing of silver behenate;
S-D is sample to detector distance (unit: mm)
λ is the wavelength of the X-ray. At 12 Kev, the wavelength is 0.1033 nm
Pixel size will be detector-dependent. (unit: mm/pixel)
Radius is the radius of silver behenate ring in the unit of pixel.